< PreviousSuperior Reliability & Performance 98 Genesis MX STM-Series Life Sciences Guide Genesis MX SPECIFICATIONS 1 Genesis MX-460 Genesis MX-480 Genesis MX-488 Genesis MX-514 Wavelength (nm) 460 ±3 480 ±3488 ±3514 ±3 FWHM Linewidth (GHz)<30 Operating ModeCW Output Power (mW) 500500500, 1000500, 1000 Power Tunability 2 (mW)50 (to 100% full rated power) Spectral Purity (%) >99 Spatial Mode TEM 00 Beam Quality (M2)<1.1 Beam Circularity 3 1.0 ±0.1 Beam Waist Diameter (mm) (FW, 1/e 2 )1.0 ±0.1 Beam Divergence (mrad) (FW, 1/e 2 ) 0.7 ±0.1 Beam Waist Location 3,4 (mm) 0.25 ±0.25 m Beam Pointing Stability 5,6 (μrad/°C) <5 Horizontal Beam Position Tolerance 6 (mm) ±<1.0 Vertical Beam Position Tolerance 6 (mm) ±<1.0 Beam Pointing Tolerance 6 (mrad) <5 Polarization Ratio >100:1 Polarization Direction Vertical, ±5° (normal to mounting plane) Noise 2 (%, rms) (10 Hz to 10 MHz)<0.1 Noise 2,6 (%, peak to peak) (10 Hz to 5 kHz)<1 Power Stability 7 (%) (pk-pk) ±<1 Warm-up Time (minutes) <10 CDRH Compliant Yes ELECTRICAL SPECIFICATIONS Operating Voltage (VAC)100 to 240 Frequency (Hz)50 to 60 Power Consumption (W)500 Operating Diode Currentn/a Maximum Diode Currentn/a Diode Voltagen/a ENVIRONMENTAL CONDITIONS Head Cooling RequirementsForced air cooled or water cooled heatsink comes included (integrated with laser); chiller for water cooled model is sold separately. Ambient Temperature Operating Non-Operating 8 10 to 40°C (50 to 104°F) -10 to 60°C (14 to 140°F) Relative Humidity 9 (%) 5 to 95, non-condensing 1 Optical parameters measured at the output plane of the laser head. Unless noted all parameters valid for the lifetime of the unit. 2 Noise specification applies at full rated power. Noise is expected to increase roughly inveresely proportionally to the output power. 3 Circularity defined as vertical diameter divided by horizontal diameter. 4 Negative value corresponds to a location inside head. 5 After 2-hour warm-up. 6 Measured at the output window. 7 Measured over 8 hrs. 8 Water cooled model needs to be completely purged of residual water before exposure to freezing temperatures. 9 Non-condensing.coherent.com 99 Life Sciences Guide Genesis MX STM-Series Genesis MX SPECIFICATIONS 1 Genesis MX 532 Genesis MX 561 Genesis MX 577 Genesis MX 590 Wavelength (nm) 532 ±3 561 ±3577 ±3590 ±3 FWHM Linewidth (GHz)<30 Operating ModeCW Output Power (mW) 500, 1000, 1500500500, 1000500, 1000 Power Tunability (mW)50 (to 100% full rated power) Spectral Purity (%) >99 Spatial Mode TEM 00 Beam Quality (M2)<1.1 Beam Circularity 3 1.0 ±0.1 Beam Waist Diameter (mm) (FW, 1/e 2 )1.0 ±0.1 Beam Divergence (mrad) (FW, 1/e 2 ) 0.7 ±0.1 Beam Waist Location 3,4 (mm) 0.25 ±0.25 m Beam Pointing Stability 5,6 (μrad/°C) <5 Horizontal Beam Position Tolerance 6 (mm) ±<1.0 Vertical Beam Position Tolerance 6 (mm) ±<1.0 Beam Pointing Tolerance 6 (mrad) <5 Polarization Ratio >100:1 Polarization Direction Vertical, ±5° (normal to mounting plane) Noise 2 (%, rms) (10 Hz to 10 MHz)<0.1 Noise 2,6 (%, peak to peak) (10 Hz to 5 kHz)<1 Power Stability 7 (%) (pk-pk) ±<1 Warm-up Time (minutes) <10 CDRH Compliant Yes ELECTRICAL SPECIFICATIONS Operating Voltage (VAC)100 to 240 Frequency (Hz)50 to 60 Power Consumption (W)500 Operating Diode Currentn/a Maximum Diode Currentn/a Diode Voltagen/a ENVIRONMENTAL CONDITIONS Head Cooling RequirementsForced air cooled or water cooled heatsink comes included (integrated with laser); chiller for water cooled model is sold separately. Ambient Temperature Operating Non-Operating 8 10 to 40°C (50 to 104°F) -10 to 60°C (14 to 140°F) Relative Humidity 9 (%) 5 to 95, non-condensing 1 Optical parameters measured at the output plane of the laser head. Unless noted all parameters valid for the lifetime of the unit. 2 Noise specification applies at full rated power. Noise is expected to increase roughly inveresely proportionally to the output power. 3 Circularity defined as vertical diameter divided by horizontal diameter. 4 Negative value corresponds to a location inside head. 5 After 2-hour warm-up. 6 Measured at the output window. 7 Measured over 8 hrs. 8 Water cooled model needs to be completely purged of residual water before exposure to freezing temperatures. 9 Non-condensing.Superior Reliability & Performance 100 Genesis MX STM-Series Life Sciences Guide Genesis MX MECHANICAL SPECIFICATIONS Dimensions (L x W x H) Laser Head 1 with Air-Cooled Heatsink (End User) Laser Head 1 with Water-Cooled Heatsink (End User) Power Supply (End User) Laser Head (OEM) Power Supply (OEM) Cables (laser head to controller) 201 x 146 x 89 mm (7.9 x 5.7 x 3.5 in.) 201 x 100 x 89 mm (7.9 x 3.9 x 3.5 in.) 361 x 229 x 180 mm (14.2 x 9.0 x 7.1 in.) n/a n/a 3 m (9.8 ft.) Weight Laser Head (End User) Power Supply (End User) Laser Head (OEM) Power Supply (OEM) Water-Cooled 4.9 kg / Air-Cooled 3.8 kg 6.0 kg n/a n/a 1 Back connector not included in laser head length dimension. MECHANICAL SPECIFICATIONS 78.74 mm (3.1 in.) 200.86 mm (7.91 in.) 219.91 mm (8.66 in.) 67.26 mm (2.65 in.) 61.6 mm (2.43 in.) 13.53 mm (0.53 in.) 10.41 mm (0.41 in.) 10.0 mm (0.39 in.) 0.0 mm (0.0 in.) 36.96 mm (1.45 in.) 53.96 mm (2.12 in.) 165.96 mm (6.53 in.) 182.96 mm (7.20 in.) Mounting Holes M5 - 0.8 X 0.38 in. Deep (4X) Mounting Holes 10-32 X 0.38 in. Deep (4X) Alignment Slot 0.251/0.252 in. Wide X 0.25 in. Long X 0.38 in. Deep Alignment Hole 0.251/0.252 in. Dia. X 0.38 in. Deep 61.6 mm (2.43 in.) 55.0 mm (2.17 in.) 45.0 mm (1.77 in.) M5 M5 M5 M5 10-32 10-32 10-32 10-32 68.2 mm (2.69 in.) 47.64 mm (1.88 in.) 89.22 mm (3.51 in.) Front View Top View Side View Rear View Genesis MX STM-Series (End-User) Air-Cooled versioncoherent.com 101 Life Sciences Guide Genesis MX STM-Series Genesis MX MECHANICAL SPECIFICATIONS Genesis MX STM-Series (End-User) Water-Cooled version 78.74 mm (3.1 in.) 61.6 mm (2.43 in.) 200.86 mm (7.91 in.) 219.91 mm (8.66 in.) 21.29 mm (0.84 in.) 13.53 mm (0.53 in.) 10.41 mm (0.41 in.) 10.0 mm (0.39 in.) 0.0 mm (0.0 in.) 36.96 mm (1.45 in.) 53.96 mm (2.12 in.) 165.96 mm (6.53 in.) 182.96 mm (7.20 in.) Mounting Holes M5 - 0.8 X 0.38 in. Deep (4X) Mounting Holes 10-32 X 0.38 in. Deep (4X) Alignment Slot 0.251/0.252 in. Wide X 0.25 in. Long X 0.38 in. Deep Alignment Hole 0.251/0.252 in. Dia. X 0.38 in. Deep 61.6 mm (2.43 in.) 55.0 mm (2.17 in.) 45.0 mm (1.77 in.) M5 M5 M5 M5 10-32 10-32 10-32 10-32 68.2 mm (2.69 in.) 47.64 mm (1.88 in.) 89.22 mm (3.51 in.) Front View Top View Side View Rear ViewSuperior Reliability & Performance 102 Genesis MX STM-Series Life Sciences Guide Genesis MX SPECIFICATIONS 1 Genesis MX 460 OEM Genesis MX 480 OEM Genesis MX 488 OEM Genesis MX 514 OEM Wavelength (nm) 460 ±3 480 ±3488 ±3514 ±3 FWHM Linewidth (GHz)<30 Operating ModeCW Output Power (mW) 500500500, 1000500, 1000 Power Tunability 2 (mW)Range is dependent on driving electronics; 50 mW to 100% full rated power with Coherent controller / power supply Spectral Purity (%) >99 Spatial Mode TEM 00 Beam Quality (M2)<1.1 Beam Circularity 3 1.0 ±0.1 Beam Waist Diameter (mm) (FW, 1/e 2 )1.0 ±0.1 Beam Divergence (mrad) (FW, 1/e 2 ) 0.7 ±0.1 Beam Waist Location 3,4 (mm) 0.25 ±0.25 m Beam Pointing Stability 5,6 (μrad/°C) <5 Horizontal Beam Position Tolerance 6 (mm) ±<1.0 Vertical Beam Position Tolerance 6 (mm) ±<1.0 Beam Pointing Tolerance 6 (mrad) <5 Polarization Ratio >100:1 Polarization Direction Horizontal (parallel to heatsink) Noise 2 (%, rms) (10 Hz to 10 MHz)<0.1 Noise 2,6 (%, peak to peak) (10 Hz to 5 kHz)<1 Power Stability 7 (%) (pk-pk) ±<1 (with Coherent controller/power supply) Warm-up Time (minutes) <10 (with Coherent controller/power supply and heatsink) CDRH Compliant No ELECTRICAL SPECIFICATIONS Operating Voltage (VAC)n/a Frequency (Hz)n/a Power Consumption (W)n/a Operating Diode Current<12.5<10<10, <12.5<10, <12.5 Maximum Diode Current<15<12<12, <15<12, <15 Diode Voltage1.5 to 2.2 ENVIRONMENTAL CONDITIONS Head Cooling RequirementsHeat sink required (available from Coherent); case temperature must be maintained at 25 ±1°C Ambient Temperature Operating Non-Operating 8 Dependent on laser head mounting and heatsinking; 10 to 40°C with Coherent heatsink -10 to 60°C (14 to 140°F) Relative Humidity 9 (%) 5 to 95, non-condensing 1 Optical parameters measured at the output plane of the laser head. Unless noted all parameters valid for the lifetime of the unit. 2 Noise specification applies at full rated power. Noise is expected to increase roughly inveresely proportionally to the output power. 3 Circularity defined as vertical diameter divided by horizontal diameter. 4 Negative value corresponds to a location inside head. 5 After 2-hour warm-up. 6 Measured at the output window. 7 Measured over 8 hrs. 8 Water cooled model needs to be completely purged of residual water before exposure to freezing temperatures. 9 Non-condensing.coherent.com 103 Life Sciences Guide Genesis MX STM-Series Genesis MX SPECIFICATIONS 1 Genesis MX 532 OEM Genesis MX 561 OEM Genesis MX 577 OEM Genesis MX 590 OEM Wavelength (nm) 532 ±3561 ±3577 ±3590 ±3 FWHM Linewidth (GHz)<30 Operating ModeCW Output Power (mW) 500, 1000, 1500500500, 1000500, 1000 Power Tunability 2 (mW)Range is dependent on driving electronics; 50 mW to 100% full rated power with Coherent controller / power supply Spectral Purity (%) >99 Spatial Mode TEM 00 Beam Quality (M2)<1.1 Beam Circularity 3 1.0 ±0.1 Beam Waist Diameter (mm) (FW, 1/e 2 )1.0 ±0.1 Beam Divergence (mrad) (FW, 1/e 2 ) 0.7 ±0.1 Beam Waist Location 3,4 (mm) 0.25 ±0.25 m Beam Pointing Stability 5,6 (μrad/°C) <5 Horizontal Beam Position Tolerance 6 (mm) ±<1.0 Vertical Beam Position Tolerance 6 (mm) ±<1.0 Beam Pointing Tolerance 6 (mrad) <5 Polarization Ratio >100:1 Polarization Direction Horizontal (parallel to heatsink) Noise 2 (%, rms) (10 Hz to 10 MHz)<0.1 Noise 2,6 (%, peak to peak) (10 Hz to 5 kHz)<1 Power Stability 7 (%) (pk-pk) ±<1 (with Coherent controller/power supply) Warm-up Time (minutes) <10 (with Coherent controller/power supply and heatsink) CDRH Compliant No ELECTRICAL SPECIFICATIONS Operating Voltage (VAC)n/a Frequency (Hz)n/a Power Consumption (W)n/a Operating Diode Current<12.5<10<10, <12.5<10, <12.5 Maximum Diode Current<15<12<12, <15<12, <15 Diode Voltage1.5 to 2.2 ENVIRONMENTAL CONDITIONS Head Cooling RequirementsHeat sink required (available from Coherent); case temperature must be maintained at 25 ±1°C Ambient Temperature Operating Non-Operating 8 Dependent on laser head mounting and heatsinking; 10 to 40°C with Coherent heatsink -10 to 60°C (14 to 140°F) Relative Humidity 9 (%) 5 to 95, non-condensing 1 Optical parameters measured at the output plane of the laser head. Unless noted all parameters valid for the lifetime of the unit. 2 Noise specification applies at full rated power. Noise is expected to increase roughly inveresely proportionally to the output power. 3 Circularity defined as vertical diameter divided by horizontal diameter. 4 Negative value corresponds to a location inside head. 5 After 2-hour warm-up. 6 Measured at the output window. 7 Measured over 8 hrs. 8 Water cooled model needs to be completely purged of residual water before exposure to freezing temperatures. 9 Non-condensing.Superior Reliability & Performance 104 Genesis MX STM-Series Life Sciences Guide MECHANICAL SPECIFICATIONS Dimensions (L x W x H) Laser Head 1 with Air-Cooled Heatsink (End User) Laser Head 1 with Water-Cooled Heatsink (End User) Power Supply (End User) Laser Head (OEM) Power Supply (OEM) Cables (laser head to controller) 201 x 146 x 89 mm (7.9 x 5.7 x 3.5 in.) 201 x 100 x 89 mm (7.9 x 3.9 x 3.5 in.) 361 x 229 x 180 mm (14.2 x 9.0 x 7.1 in.) n/a n/a 3 m (9.8 ft.) Weight Laser Head (End User) Power Supply (End User) Laser Head (OEM) Power Supply (OEM) Water-Cooled 4.9 kg / Air-Cooled 3.8 kg 6.0 kg n/a n/a 1 Back connector not included in laser head length dimension. MECHANICAL SPECIFICATIONS Slot 3.2 +0.025/-0.0 mm (0.126 + 0.001/-0.0 in.) Wide X 2.03 mm (0.08 in.) Long X 5.08 mm (0.20 in.) Deep Mounting Holes M3 X 12.2 mm (0.48 in.) Deep (4X) 3.2 +0.025/-0.0 mm (0.126 + 0.001/-0.0 in.) DIA X 5.08 mm (0.20 in.) Deep 5.46 mm (0.22 in.) 120.65 mm (4.75 in.) 111.1 mm (4.37 in.) 34.55 mm (1.36 in.) 4.78 mm (0.19 in.) 61 mm (2.4 in.) 34 mm (1.34 in.) 17 mm (0.67 in.) 22 mm (0.87 in.) 44 mm (1.73 in.) 26 mm (1.02 in.) 29.78 mm (1.17 in.) 28.15 mm (1.11 in.) 18.8 mm (0.74 in.) 18.8 mm (0.74 in.) 24.89 mm (0.98 in.) 6.1 mm (0.24 in.) 5.08 mm (0.2 in.) 45.21 mm (1.78 in.) 6.35 mm (0.25 in.) 6.35 mm (0.25 in.) 6.35 mm (0.25 in.) 5.08 mm (0.2 in.) 16.26 mm (0.64 in.) Spacer Detail Clamping Bar Detail 3.175 mm (0.125 in.) 1.52 mm (0.06 in.) 12.95 mm (0.51 in.) 4.06 mm (0.16 in.) 86.87 mm (3.42 in.) 89.92 mm (3.54 in.) 112.78 mm (4.44 in.) 112.78 mm (4.44 in.) 120.65 mm (4.75 in.) 11.43 mm (0.45 in.) 53.59 mm (2.11 in.) 12.7 mm (0.5 in.) 15.24 mm (0.6 in.) 8-32 UNC-2B X 0.50 Deep (4X) 0.125 Dia. Dowel Pin (2X) 0.188 Dia. Thru (2X) This surface must be flat within 0.001 in. and have 32 µm finish or better within TEC mounting areas. Base Plate Requirements Front View Top View Rear View Side View Genesis MX STM-Series (OEM) Water-cooled versioncoherent.com 105 Life Sciences Guide Genesis MX STM-Series MECHANICAL SPECIFICATIONS 16.0 mm (0.63 in.) 16.0 mm (0.63 in.) 15.24 mm (0.6 in.) 127 mm (5.0 in.) 22.43 mm (0.88 in.) 18.34 mm (0.72 in.) 77.65 mm (3.06 in.) 42.15 mm (1.66 in.) 111.76 mm (4.4 in.) 111.0 mm (4.37 in.) 15.24 mm (0.60 in.) M36-32 6-32 6-32 6-32 M3 M3 M3 M3 6-326-32 6-32 M3M3 M3 0 mm (0.0 in.) RS-232D Serial Port USB Interlock 24V In 127 mm (5.0 in.) 257.15 mm (10.12 in.) 0 mm (0.0 in.) 0 mm (0.0 in.) 52.37 mm (2.06 in.) 68.58 mm (2.70 in.) 243.31 mm (9.58 in.) 204.77 mm (8.06 in.) 188.57 mm (7.42 in.) 52.37 mm(2.06 in.)68.58 mm(2.70 in.) 204.77 mm(8.06 in.)188.57 mm(7.42 in.) 111 mm (4.37 in.) 111.76 mm (4.40 in.) Mounting Holes 10X On Bottom As Indicated Mounting Holes (both sides) GND Diode Out Laser Control TEC Ext. Fan Front View Rear View Top View Side View Bottom View Genesis MX STM-Series Low Current OEM ControllerSuperior Reliability & Performance 106 Coherent follows a policy of continuous product improvement. Specifications are subject to change without notice. Coherent’s scientific and industrial lasers are certified to comply with the Federal Regulations (21 CFR Subchapter J) as administered by the Center for Devices and Radiological Health on all systems ordered for shipment after August 2, 1976. Coherent offers a limited warranty for all Genesis MX SLM lasers. For full details of this warranty coverage, please refer to the Service section at www.Coherent.com or contact your local Sales or Service Representative. ISO 9001 Registered Genesis MX STM-Series Life Sciences Guide Genesis MX MECHANICAL SPECIFICATIONS Genesis MX STM-Series Benchtop Power Supply 159.8 mm (6.29 in.) 19.8 mm (0.78 in.) 361.1 mm (14.22 in.) 228.9 mm (9.01 in.) Side ViewFront Viewcoherent.com 107 Genesis MX SLM-Series Single Frequency Visible OEM and End-User OPS Laser Systems Applications like Flow Cytometry, Particle Counting, DNA Sequencing and Microscopy are enable by low noise, visible true CW lasers. The Genesis MX SLM-Series provides up to 1 W of visible laser light from either OEM or CDRH-compliant end-user systems. Based on Coherent’s unique Optically Pumped Semiconductor Laser (OPSL) technology, the Genesis MX SLM-Series features single frequency operation for the most demanding applications. This, combined with stable beam parameters across output powers, a diffraction-limited beam, low noise and high stability, provides unparalleled laser perfor- mance in a convenient package. Genesis MX SLM-Series is the perfect match for customers in need of the highest performing CW laser technology for research and instru- mentation in life science and biological applications. FEATURES & BENEFITS • All Genesis MX advantages with single-frequency output • OEM or end-user versions • Air- or water-cooled solutions APPLICATIONS • Flow Cytometry • Particle Counting • DNA Sequencing • Microscopy Genesis MXNext >